5 matches found
CVE-2018-19012
The CVE-2018-19012 issue affects Dräger Infinity Delta, Delta XL, Kappa, and Infinity Explorer C700 (all versions). The root cause is improper privilege management in a dialog that allows breaking out of kiosk mode, enabling an attacker to reach the underlying OS and take control of the system. T...
CVE-2018-19010
CVE-2018-19010 affects Dräger Infinity Delta family (Infinity Delta, Delta XL, Kappa, Infinity Explorer C700 all versions). Root cause: improper input validation (CWE-20) via a malformed network packet that causes the monitor to reboot. Impact: attacker-supplied malformed packets could repeatedly...
CVE-2018-19014
CVE-2018-19014 affects Dräger Infinity Delta, Delta XL, Kappa, and Infinity Explorer C700 (all versions). The root issue is information exposure through log files: log data is accessible over an unauthenticated network connection, enabling an attacker to gain insights into internals of the patien...
CVE-2019-25716
The CVE affects Dräger Infinity Delta, Delta XL, and Kappa patient monitors. A denial-of-service flaw lets an attacker send malformed network packets to reboot the device, disrupting monitoring until it reverts to default configuration and loses network connectivity. Documented metrics show CVSSv...
CVE-2019-25717
CVE-2019-25717 affects Dräger Infinity Delta, Delta XL, and Kappa patient monitors. The vulnerability allows unauthenticated attackers over a network to access exposed log files, exposing device internals, location data, and wired network configuration details. No root cause or remediation detail...